Environmental Conditions For Installation And Use - B.A. International BASE900 Manuel D'utilisation

Ba alginate mixer
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

6
7. The Mixing bowl is replaceable. Renew if damage or difficulty in closure is experienced.
8. There are two retention bolts (for immobilisation) at the base of the unit, these should
be removed before use and replaced if the unit is moved or sent for repair, (Fig.1) this
prevents unnecessary damage during transit.

ENVIRONMENTAL CONDITIONS FOR INSTALLATION AND USE :

1. Conditions for installations and use:
1) For indoor use only.
2) Operating ambient temperature is between 2°C ~ 40°C (35.6 ~ 104°F).
3) Operating altitude is below 2000 metres and atmospheric pressures between 900 ~
1600 millibars.
4) Relative Humidity is lower than 80%.
5) Operating area should be free of dust, acids or solvent vapours.
6) The required electric voltage: AC110V (+/- 15V); AC220 (+/- 20V).
2. Installation:
1) The BASE900 BA Alginate Mixer should be positioned on a firm, flat, vibration free
surface.
2) Never install near to inflammable substances.
3) Keep away from water or moisture to avoid electric shock or mechanical problems.
4) Do not place anything on the cover or control panel to avoid any damage or spillage.
3. Before/After Transportation:
1) Before installing the BASE900 BA Alginate Mixer remove the
two bolts fixed at thebbottom and follow the instruction (Fig.1)
to store them safely.
2) If the mixer needs to be moved or sent for repair, replace the
bolts to immobilize the motor to prevent possible malfunction
or damage.
4. Power:
1) Check the power switch on the back of the unit is off, and then connect the power
cord to the voltage adaptor. (Fig. 2)
2) Plug into a standard AC outlet, with co-axial cable of AC 220V 50/60Hz with a 3Amp
fuse or AC110V 50/60Hz with a 5Amp fuse. In order to avoid electric shock, the electric
power cord shall be earthed.
GEFAHR:
- Check to see the BASE900 BA Alginate Mixer's voltage requirement matches
your power source. Damage to the BASE900 BA Alginate Mixer, electric shock, or
fire may occur if voltages are incompatible.
- Keep any potentially dangerous substances a minimum of 30cm (1ft) radius from
the BASE900 BA Alginate Mixer.
- Ensure that hand and body are dry when touching power socket, main switch or
operating the equipment.