9. Post-use Maintenance

After each patient maintain the product as follows.
WARNING
• I mmediately after a treatment, the handpiece should be
cleaned, lubricated and sterilized. Failure to properly maintain
the handpiece may lead to overheating, causing burn injuries or
product failure.
CAUTION
• D o not use the following fluids to wipe, immerse or clean the
product; strong/super acid water, strong acid/alkaline chemicals,
chlorine-containing solutions, solvents such as benzine or thinner.
9-1 Preparation
1) Wear a eye protection, a mask and gloves to avoid infection.
2) Disconnect the handpiece from the motor.
4
3) Remove the motor from the hose.
9-2 Cleaning
Manual Cleaning (External Surface)
1) Remove dirt and debris on the motor surface using a brush under
running water (at 38°C or less). Do not use a wire brush.
2) Wipe off the moisture on the surface using absorbent cloth.
Automatic Cleaning (External and Internal Surfaces)
Use a thermo-disinfector in accordance with ISO 15883 (EN ISO 15883).
Refer to the thermo-disinfector manual.
CAUTION
• A fter washing with thermo-disinfector and prior to lubrication,
dry the product until all internal moisture is thoroughly removed.
Thermo-disinfector moisture remaining inside the product could
reduce the effect of lubrication and could cause corrosion inside
of the product.