Operation Manual.
• D o not allow any impact on to the product. Do not drop the
product.
• O perators and all others in the area must wear eye protection and
a mask when operating this motor.
• S hould the product function abnormally, cease operation
immediately and contact your Authorized B.A. International Dealer.
• D o not use high acid water or sterilizing solutions to wipe, immerse
or clean the product.
• T he products are delivered in a non-sterile condition and must be
autoclaved prior to use.
• P erform regular function and maintenance checks.
• I f the product is not used for a long period check it is functioning
correctly before using on a patient.
• T o avoid clinical downtime it is recommended that a spare be kept
on hand in case of a breakdown during surgery.
2

3. Accessory List

Part Name
Quantity
Lubrication Oil
1
4. Setting of Air & Water Supply Pressure
Measure the supply pressure at the handpiece / hose connection point
and set the pressure to the value specified on the specification table.
(Fig. 1)
For Multi Gauge information refer to Option Parts List.
WARNING
• D o not exceed the optimum pressure specified on the specification
table.
CAUTION
• D o not use air contaminated by dust, moisture and oil.