MAINTENANCE

7.1

Error message

When ST3100M is in pH mode:
Error 0
Memory access error
Error 1
Self-diagnosis failed
Error 2
Measured values out of
range
Error 3
Measured buffer temperature
out of range (<5 or >40 ° C)
Error 4
Offset out of range
offset > 60mV or < - 60 mV
Error 5
Slope out of range
Error 6
Meter cannot recognize the
buffer
Error 9
The current data set has
already been stored once
Error 10
The sample temperature is
out of range
When ST3100M is in conductivity mode:
Error 0
Memory access error
Error 1
Self-diagnosis failed
Error 2
Measured values out of range
C:
> 199.9 mS/cm
TDS: < 0.1 mg/L or > 199.9 g/L
SAL: > 100.0 psu
Error 3
Measured standard temperature out
of range (5 ... 35 ° C)
Error 4
Measuring temperature out of range
(0 ... 100 ° C)
Error 9
The current data set has already
been stored once
If an error occurs, the meter will also beep 3 times to alert.
For further technical support please contact Ohaus. (US please contact 1-800-672-7722).
STARTER 3100M Bench Meter
Reset to factory settings
Repeat the self-diagnosis procedure and make
sure that you finish pressing all five keys within two
minutes
Check if the electrode is properly connected and
placed in the sample solution.
Keep the pH buffer temperature within the range
for calibration
Make sure the pH buffer is correct and fresh;
Clean or replace the pH electrode.
Make sure the buffer is correct and fresh; Clean or
replace the pH electrode.
Make sure the buffer is correct and fresh; check if
the buffer has not been used more than once.
An endpoint reading can only be stored once.
Perform a new measurement to store.
Check the sample temperature, the temperature
sensor.
Reset to factory settings
Repeat the self-diagnosis procedure and
make sure that you finish pressing all five
keys within two minutes.
Make if the electrode is properly connected
and placed in the sample solution
Keep the standard temperature within the
range for calibration
Check if the electrode is properly connected
and keep the sample temperature within the
range
An endpoint reading can only be stored
once. Perform a new measurement to store.