Performing 2-Point Calibration - OHAUS STARTER 3100M Manuel D'instruction

Analyseur de table multi paramètre
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN-12
Press
and
appear on the top of the screen,
The meter reaches endpoint automatically according to the preselected auto-endpoint
mode, the calibration point pH value (e.g. 7.00) with the temperature display on the screen.
The 1-point calibration is finished; There are now 3 options (OHAUS recommends conducting at
least a 2 point calibration);
1.
Press
2.
Press
shown on the display for 3 seconds then return to the measurement screen.
3.
Press
3.2.2

Performing 2-point calibration

Perform 1-point calibration as described above.
Rinse the pH electrode with pure water and wipe off with tissue.
Place the electrode in the next calibration buffer, stir and wait, then press
displays on the bottom left of the screen and "
and
appear,
The meter reaches endpoint according to the endpoint mode, the calibration point pH value
(e.g. 4.01) with the temperature display on the screen.
The 2-point calibration is finished. There are 3 options:
1.
Press
2.
Press
shown on the display for 3 seconds then return to the measurement screen.
3.
Press
Note: To perform a 3 point calibration, follow the instructions for performing a 2 point calibration
using a third buffer.
"
1" displays on the bottom left of the screen and "
Cal
to do the 2-point calibration.
to store the 1-point calibration and exit, the offset and the slope are
to reject the calibration, return to the measurement screen.
is blinking during calibration.
to do the 3-point calibration.
to store the 2-point calibration and exit, the offset and slope are
to reject the calibration, return to the measurement screen.
STARTER 3100M Bench Meter
is blinking during calibration.
" is blinking. On the top of the screen
Cal
" is blinking.
Cal
, "
2"
Cal