Test Of The Direction Of Phase Rotation - Chauvin Arnoux C.A 6030 Notice De Fonctionnement

Masquer les pouces Voir aussi pour C.A 6030:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
4.2.2.6 Warnings or error reports
Preliminary remark : The complete list of coded errors is given in § 7.
Display - Indication
> 80°C
HOt
Er07
U
FN
Er11
Er12
Ta > 500ms
Er08
n PE
Er19
Er23
I
> 1,06 I
∆ ∆ ∆ ∆ ∆ N
A
Press the
key to exit from the error conditions.
TEST
4.3

TEST OF THE DIRECTION OF PHASE ROTATION

4.3.1 Description of the function
This function is used to check the order of the phases in a three-phase network using 2 cables ("2-
wire" method, sequential measurement) or 3 cables ("3-wire" method, static measurement).
The "2-wire" measurement, which is simple, is well suited to tests on distribution frames.
The "3-wire" measurement is preferable in the case of tests on motors, generators, etc.
Note: The phase order is indicated after the acquisition of one reference period (time) then one
measurement period (time).
Remark - Possible cause
The temperature of the instrument is too high: the
measurement is stopped. Pressing the
effect until the temperature of the instrument is again below
60°C; another measurement can then be started.
U
< U
(see § 3.2).
FN
L
The test is broken off (no non-tripping test)
Unplanned tripping of the RCDr (measurement current +
leakage current too high): the measurement is broken off.
The RCD failed to trip in the 2
RCD defective in pulse mode.
The RCD tested tripped during the loop measurement ;
there is a reversal of the N and PE terminals.
The type of test chosen is incompatible with the assigned
current of the RCD (see the tables of § 4.2.2.2)
The RCD failed to trip although the current was greater
mA
than I
: RCD defective in sweep mode.
∆N
68
key has no
TEST
nd
test :

Publicité

Table des Matières
loading

Table des Matières