Where the various indicators represente, respectively :
1
position of the phase conductor
2
detected voltage probe
3
measurement smoothed for display
4
audible buzzer activated
5
call in progress (serial link)
6
battery charge level
7
standby switching function deactivated
8
level of memory use
9
reading/recording in memory
10
secondary display unit A2
11
"test" number for storage
12
"object" number for storage
13
main display unit A1
14
half-wave signal for the RCD test
1.1

ENVIRONMENTAL CONDITIONS

Temperature:
Service conditions: -10 to +55°C - storage and transport (without batteries): -40 to +70°C.
%RH
Service conditions: 85% max. - storage and transport (without batteries): 90% max.
(without condensation):
Tightness:
IP54 as per standard NF EN 60 529.
1.2

COMPLIANCE WITH STANDARDS AND SAFETY

1.2.1

COMPLIANCE WITH STANDARDS

The instrument complies with the following standards :
- EN 61010-1 (Ed. 2001)
- NF EN 61557 (Ed. 97: parts 1 to 7, ed. 2001: part 10)
1.2.2
SAFETY
The instrument complies with the requirements of standards EN 61010-1 and EN 61557, i.e.:
-
service voltage: 550 V,
-
measurement category: III with double insulation,
-
level of pollution: 2.
-
Never use on DC or AC networks exceeding 550V with respect to ground.
This instrument is designed for indoor use at an altitude < 2000m
Never open the housing of this instrument until it has been disconnected from any source of
electricity.
Before making any measurement, make sure that the positions of the cords and the setting of the
switch are correct.
Never connect to the circuit to be measured if the housing of the instrument is not closed.
The protection provided by the instrument may be compromised if it is used in a manner other than
that specified
1.2.3

ELECTROMAGNETIC COMPATIBILITY

CE instrument, in conformity with product standard EN 61326-1 (ed. 97) + A1 (ed. 98):
Emissions: Requirements on class B equipment.
Immunity:
Requirements on equipment used in discontinuous operation on industrial sites.
1.3

POWER SUPPLY

-
Power supply :
-
Battery life :
15
full-wave signal for the testing of RCDs
16
measurement without triggering of differentials (signalling level)
17
measurement with triggering of differentials (power level)
18
RCD test current reading in "pulse" mode
19
test of residual current devices in "pulse" mode
20
test of residual current devices in "sweep" mode
21
secondary function activated
22
type of quantity displayed
23
selective measurement
24
alarm function activated or display of an alarm threshold
25
"WARNING" indicator (if it appears, refer to the manual)
26
reference voltage
27
time to tripping of the RCD
28
compensation of measuring cables activated
6 LR6 1.5V alkaline batteries; they can be replaced by rechargeable batteries
having a capacity of at least 1,800mAh.
30 hours or approximately
-
10, 000 loop measurements
-
30, 000 voltage or current measurements during 5 seconds.