9.2. Disinfection and sterilization procedure

#
Operation
Preparation at the
1
point of use prior
to processing
Preparation for
decontamination/
2
preparation before
cleaning
Cleaning:
3
Automated
4
Cleaning: Manual
5
Disinfection
6
Drying
Maintenance,
inspection and
7
testing of the
accessories
8
Packaging
9
Sterilization
Instructions
No particular
requirements.
No particular
requirements.
The accessories are
not intended for
automated cleaning.
Clean the accessories
with an appropriate
brush or towel soaked
in a disinfectant
solution.
Soak the accessories
in a disinfectant
solution that contains
a proteolytic enzyme if
possible. Rinse
accessories
thoroughly in running
water.
No particular
requirements.
No particular
requirements.
Pack the devices in
sterilization pouches.
- Steam sterilization at
135°C (275°F) for 10
minutes in tabletop,
N-type autoclave.
- Drying time after
sterilization – 30
minutes.
14
Details and Warnings
- The file clip should be pressed
and released several times during
cleaning to assure all debris is
removed.
- After cleaning, there should be no
visible debris on the accessories.
- Follow manufacturer's instructions
on the disinfectant (concentration,
immersion time, etc.).
- Do not use disinfectant containing
aldehyde, phenol or any products
which may damage the items.
- Check the expiration date of the
pouch given by the manufacturer
to determine the shelf life.
- Use packaging which is resistant
up to a temperature of 141°C
(286°F).
- Follow maintenance and
operation procedures of the
autoclave provided by the
manufacturer.
- The only sterilization parameters
to be used are those that have
been provided in this Manual.