Dental Material Recommendations; Hygiene; Before Patient Treatment; During Patient Treatment - B.A. International Ultimate 1400 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 30
10
1. Prior to using the Ultimate 1400, attached the rubber eye-protector on to the light guide and use
protective, light-filtering eyewear (ex. glasses).
2. Only use the Ultimate 1400 after it has completed an adequate hygiene procedure.
3. Protect the Ultimate 1400 by using single-use, protective, plastic barrier-sleeves.
4. Never place the Ultimate 1400 handpiece, light guide, or depleted battery packs on the recharging
stations before they are cleaned and disinfected.

4. Dental Material Recommendations

In general, the Ultimate 1400 has been optimized to achieve a 10 second curing time with the latest generation dental
materials which cure with blue and violet light (400 nm to 480 nm). These materials include restoratives, bonding agents
and adhesives, bases, liners, fissure sealants, some temporary material, and luting materials for brackets as well as
dental-lab materials such as ceramic inlays. However, manufacturer's instructions regarding specific light curing dental
materials should be followed.

5. Hygiene

5.1. Before Patient Treatment

Check that the light output via the radiometer and inspect the light guide to assure there are no obstructions.
Check that the eye-protector is installed on the light guide properly.
Insert the Ultimate 1400 into a protective, single-use, barrier sleeve for additional protection if desired. Be certain
the sleeve is tightly fitted to the Ultimate 1400 to avoid treatment complications.
Outfit dental personnel and patient with appropriate protective eyewear (light filtering, infection protection).
Consider patient's health needs, potential light sensitivity and determine the appropriate precautionary actions.
Place the Ultimate 1400 handpiece in a convenient place for use during treatment.

5.2. During Patient Treatment

Never place the Ultimate 1400 handpiece, light guide, or depleted battery packs on the recharging stations before
they are cleaned and disinfected.
Do not use the Ultimate 1400 as a tissue retractor to manipulate the tongue and/or cheeks.
1. Prior to treatment, attached the rubber eye-protector on to the light guide and use protective, light-
filtering eyewear (ex. glasses).
2. Only use the Ultimate 1400 after it has completed an adequate hygiene procedure.
3. Protect the Ultimate 1400 by using single-use, protective, plastic barrier-sleeves.
4. Never place the Ultimate 1400 handpiece, light guide, or depleted battery packs on the recharging
stations before they are cleaned and disinfected.

5.3. After Patient Treatment

Immediately remove the protective, single-use, barrier sleeve and dispose of according to local regulations and
manufacturer's instructions. During removal, be careful not to contaminate the Ultimate 1400.
Discard examination gloves according to local regulations and disinfect hands with a bactericidal, virucidal, and
fungicidal hand disinfectant solution according to local regulations. Use according to disinfectant solution manu-
facturer's instructions for Use. Use a new, clean pair of examination gloves for cleaning and disinfection of the
Ultimate 1400.
Remove eye-protector. Clean and disinfection this separately.
Remove bioburden from the Ultimate 1400.
Disinfect with alcohol-based fluids by wiping the complete exterior of the Ultimate 1400 components. Use a single-
use cloth for wiping the unit. Do not submerge and be careful to avoid getting the battery pack contacts wet. Do not
allow fluids to penetrate into the handpiece.
The battery pack can be removed after the Ultimate 1400 exterior has been cleaned and disinfected.
The light guide and the eye-protector shield can be autoclaved using a validated sterilization process at a maximum
temperature of 134 °C and for a time of up to 20 min. Perform sterilization according to EN 17665-1:2006 and EN
556-1:2001 at 134 °C for at least 3 minutes and use steam sterilizers that comply with the requirements of EN
13060:2004+A2:2010, class B or S.
After cleaning and disinfection, place the Ultimate 1400 and battery pack back on the recharging station.
Ultimate 1400.indd 10
30.01.15 16:14

Publicité

Table des Matières
loading

Table des Matières