Specifications
Warning: The technical specifications of the mask are provided for your healthcare
professional to determine if it is compatible with your CPAP or bi-level therapy device. If
used outside these specifications, or if used with incompatible devices, the mask may be
uncomfortable, the seal of the mask may not be effective, optimum therapy may not be
achieved, and leak, or variation in the rate of leak, may affect device function.
Pressure Flow Curve
90
80
70
60
50
40
30
20
10
0
0
5
Resistance with Anti-Asphyxia Valve Closed to Atmosphere
Drop in Pressure at
All sizes
Inspiratory and Expiratory Resistance with Anti-Asphyxia Valve Open to Atmosphere
Inspiratory Resistance
Expiratory Resistance
Deadspace Volume
P
224 ml
S
231 ml
M
242 ml
L
253 ml
Sound Levels
A-weighted Sound Power Level
A-weighted Sound Pressure Level at 1 m
Disposal
Dispose of in accordance with local regulations.

Storage Conditions

Temperature: -4º to 140º F (-20º to +60º C)
Relative Humidity: 15% to 95% non-condensing
6
Pressure Flow Curve
Amara Full Face Mask
10
15
Pressure cm H
O
2
50 SLPM
0.6 cm
H
O
2
20
25
30
100 SLPM
1.5 cm
H
2
0.6 cm
H
O
2
0.7 cm
H
O
2
28 dBA
20 dBA