The device complies with all applicable and required standards for electromagnetic
interference.
It does not normally affect nearby equipment and devices.
It is not normally affected by nearby equipment and devices.
It is not safe to operate the scale in the presence of high-frequency surgical
equipment.
However, it is good practice to avoid using the device in extremely close proximity to
other equipment.

Emissions and immunity information

The Welch Allyn Masted Scale is intended for use in the electromagnetic environment specified below. The customer or
user of the Welch Allyn Masted Scale should assure that it is used in such an environment.
Emissions test
Radiated emissions
CISPR 11
Conducted emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
The Welch Allyn Masted Scale is intended for use in the electromagnetic environment specified below. The customer or
the user of the Welch Allyn Masted Scale should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Electromagnetic emissions
Compliance
Electromagnetic environment - guidance
Group 1 /Class B
The Welch Allyn Masted Scale uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Class B
The Welch Allyn Masted Scale is suitable for use in all establishments
other than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic
purposes.
Class A
Complies
Electromagnetic immunity
IEC 60601 test level
Compliance level
±8 kV contact
±8 kV contact
±15 kV air
±15 kV air
±2 kV for power supply
±2 kV for power supply
lines
lines
±1 kV for input/output
±1 kV for input/output
lines
lines
±1 kV differential mode ±1 kV differential mode Mains power quality should be that of a typical
Welch Allyn Masted Scales
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
commercial or hospital environment.