6. Commissioning, operation
6.4.2.3 Configuring data logger and starting measurement
The GDI-100-D can be configured for logging tasks either directly on the instrument or via
the "myWIKA Device" app. For this, click on the "Upload" button in the category "Logger"
and set the desired parameters.
Interval: Calculates a mean value over the duration of the set interval. If 0 is entered,
the measured values are stored in the instrument with the set sampling rate.
Duration: Defines the duration of the measurement.
Start time: Defines the start time of the measurement.
Upload: The settings are transmitted to the GDI-100-D.
Start now: Starts measurement immediately.
Stop: Stops measurement immediately.
Preview: Enables preliminary viewing of the measured value.
WIKA operating instructions, model GDI-100-D
EN
29