Error Messages And Troubleshooting - Omron M4 Intelli IT Mode D'emploi

Tensiomètre brassard
Masquer les pouces Voir aussi pour M4 Intelli IT:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 12

3. Error Messages and Troubleshooting

If any of the below problems occur during measurement, check to make sure that no other electrical device is within 30 cm. If the problem persists,
please refer to the table below.
Display/Problem
Possible Cause
The [START/STOP] button was pressed
while the arm cuff is not applied.
Air plug is not completely plugged into
appears or the arm
the monitor.
cuff does not inflate.
The arm cuff is not applied correctly.
Air is leaking from the arm cuff.
You move or talk during a measurement
appears or a
and the arm cuff does not inflate
measurement
sufficiently.
cannot be
Due to the systolic pressure is above
completed after the
210 mmHg, a measurement cannot be
arm cuff inflates.
taken.
The arm cuff is inflated exceeding the
appears
maximum allowable pressure.
You move or talk during a measurement.
appears
Vibrations disrupt a measurement.
The pulse rate is not detected correctly.
appears
/
/
appears
does not flash during
a measurement.
Solution
Press the [START/STOP] button again to turn the monitor off.
After inserting the air plug securely and applying the arm cuff
correctly, press the [START/STOP] button.
Insert the air plug securely.
Apply the arm cuff correctly, then take another measurement.
Refer to section 7 of instruction manual
Replace the arm cuff to the new one. Refer to section 14 of
instruction manual
.
Remain still and do not talk during a measurement. If "E2"
appears repeatedly, inflate the arm cuff manually until the
systolic pressure is 30 to 40 mmHg above your previous
readings. Refer to section 13 of instruction manual
Do not touch the arm cuff and/or bend the air tube while taking
a measurement. If inflating the arm cuff manually, refer to
section 13 of instruction manual
Remain still and do not talk during a measurement.
Apply the arm cuff correctly, then take another measurement.
Refer to section 7 of instruction manual
correctly during a measurement.
If the "
" symbol continues to appear, we recommend you to
consult with your physician.
EN
.
.
.
. Remain still and sit
EN4

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Hem-7155t-ebkX4 smartHem-7155t-esl

Table des Matières