SECTION 4—TYPICAL PRODUCT PARAMETERS
Tubing:
Relative Humidity:
Time of Operation:
Recommended Storage and
Shipping Temperature:
LABEL
SYMBOL
O
2
LABEL
SYMBOL
Platinum™
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
2.1 m (7 ft) cannula with a maximum 15.2 m
(50 ft) of Crush-Proof Tubing (DO NOT
pinch).
IRC9LXO2AWQ
Recommended use up to 15.2 M (50 ft) high
flow tubing with high flow cannula at all flow
rates.
20 to 60%
Up to 24 hours per day.
-29°C to 65°C (-20°F to 150°F) at 15-95%
relative humidity.
SENSO
2
STATUS
SYSTEM OKAY
O
over 85% (± 2%)
2
O
Between 73% (± 3%)
2
to 85% (±2%)
SYSTEM FAILURE
O
Below 73% (±3%)
2
LX INDICATORS
STATUS
SYSTEM OKAY
SYSTEM FAILURE
INDICATORS
INDICATOR LIGHTS (LED)
GREEN Indicator Light
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician.
RED Indicator Light
Refer to Troubleshooting Guide
on page 29.
INDICATOR LIGHTS (LED)
GREEN Indicator Light
RED Indicator light
Continuous Audible Alarm
Sieve-GARD™ Compressor
Shutdown
Call a qualified technician
16
Part No 1141491

Publicité