SECTION 7—TROUBLESHOOTING GUIDE
SYMPTOM:
YELLOW or RED
light illuminated.
Alarm:
Continuous
On red light only.
* Only applies to
SensO
models.
2
GREEN light
YELLOW Light
Flashing.
Alarm:
RAPID
Beep....Beep...
Beep...Beep
Platinum™
PROBABLE CAUSE:
1. Low oxygen purity.*
2. Kinked or blocked
tubing, cannula or
humidifier.*
3. Flowmeter set at 0.5
L/min. or less.*
4. Unit overheating due
to blocked air intake.
5. Insufficient power at
outlet.
6. Internal repairs
required.
1. Internal repairs
required.
Potential
Obstruction Alert
1. Possible internal
obstruction in the
oxygen path. Kinked or
blocked tubing, cannula
or humidifier.
2. Flowmeter set at 0.5
L/min. or less.
30
SOLUTION:
1. Clean or Replace filters.
2. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power OFF
for 60 seconds and then
turn power back ON.
3. Check flowmeter is set to
1.0 L/min. or more. (Refer
to Recommended
Guidelines for Optimum
Performance on page 18.
4.a. Remove and clean
cabinet filters.
4b. Move concentrator at
least 7.5 cm (3 in) from
walls, draperies, and
furniture.
5. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
6. Call Service Provider.
1. Call Service Provider.
1. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turnpower OFF
for 60 seconds and then
turn power back ON.
2. Reset flowmeter to
prescribed flowrate.
Flowrates less than1L/min.
are not recommended.
NOTE: The use of the
pediatric flowmeter will
deactivate the Potential
Obstruction Alert.
Part No 1141491