Problem
The charging light does
not start flashing green
when I connect the USB
cable to the computer.
The BlueTouch Pain
Relief Patch becomes
warm during use.
Model
Rated voltage (V)
Rated frequency (Hz)
Rated input (W)
Class
Classification
Ingress of water
Max. daily dose (J/cm²)
Battery type
Operation conditions
Temperature
Relative humidity
Storage conditions
Temperature
Relative humidity
Possible cause
The micro USB plug is not
inserted into the micro
USB socket properly or the
standard USB plug is not
inserted into the USB port of
your computer properly.
The standard USB plug is
not inserted or not inserted
properly into the USB port of
the computer.
The battery of the laptop into
which you have plugged the
USB cord is running low.
This is normal.
Specifications
Solution
Insert the micro USB plug into the micro
USB socket properly and check the
connection to the computer. Make sure
the computer is switched on.
Insert the standard USB plug into the USB
port of the computer properly.
Connect the laptop to the mains.
No action required.
PR3092
100-240
50-60
15
Medical device Class IIa
Risk Group 1
IP22
< 90 if treatment schedule is followed
Li-Ion
from +10°C to +35°C
from 30% to 90%
from -20°C to +35°C
from 30% to 90% (no condensation)
ENGLISH