Publicité

Les langues disponibles

Les langues disponibles

Liens rapides

D G F r
6002-9200-2011-000 29200-2016 V100 11 ⁄ 2015
1110
PREMASGARD
® 
D
Bedienungs- und Montageanleitung
Druck- und Differenzdruckmessumformer (± 3 %),
incl. Anschluss-Set, Kompaktform, einstellbar, kalibrierfähig,
mit Mehrbereichsumschaltung und aktivem Ausgang
G
Operating Instructions, Mounting & Installation
Pressure and differential pressure measuring transducers
(± 3 %), including connection set, compact form, adjustable,
calibratable, with multi-range switching and active output
F
Notice d'instruction
Convertisseur de pression et de pression différentielle (± 3 %),
y compris kit de raccordement, forme compacte, réglable,
étalonnable, avec commutation multi-gamme et sortie active
r
Руководство по монтажу и обслуживанию
Преобразователь давления измерительный и преобразователь
давления измерительный дифференциальный (±3 %), вкл.
комплект соединительных деталей, компактное исполнение,
настраиваемый, калибруемый, с переключением между
несколькими диапазонами и активным выходом
S+S REGELTECHNIK GMBH
PIRNAER STRASSE 20
90411 NÜRNBERG ⁄ GERMANY
FON +49 (0) 911 ⁄ 5 19 47- 0
FAX +49 (0) 911 ⁄ 5 19 47- 70
mail@SplusS.de
www.SplusS.de
Herzlichen Glückwunsch!
Sie haben ein deutsches Qualitätsprodukt erworben.
Congratulations!
You have bought a German quality product.
Félicitations !
Vous avez fait l'acquisition d'un produit allemand de qualité.
Примите наши поздравления !
Вы приобрели качественный продукт, изготовленный в Германии.
®
PREMASGARD
1110
®
PREMASGARD
1110
mit Display
with display
avec écran
с дисплеем

Publicité

Table des Matières
loading

Sommaire des Matières pour S+S Regeltechnik PREMASGARD 1110

  • Page 1 Display настраиваемый, калибруемый, с переключением между with display несколькими диапазонами и активным выходом avec écran с дисплеем S+S REGELTECHNIK GMBH Herzlichen Glückwunsch! Sie haben ein deutsches Qualitätsprodukt erworben. PIRNAER STRASSE 20 Congratulations! 90411 NÜRNBERG ⁄ GERMANY You have bought a German quality product.
  • Page 2 D G F r 1110 PREMASGARD ®  Maßzeichnung PREMASGARD ® 1110 ohne Display M16x1.5 without display sin écran без дисплея ø4 ø6 mit Display M16x1.5 M16x1.5 with display avec écran с дисплеем ø4 ø6 ø4 ø6...
  • Page 3: Technische Daten

    ® PREMASGARD 1110 Rev. 2016 - V11 DE Qualitätsprodukt für HKL-Bereich, Genauigkeit ± 3 % Die kalibrierfähigen, kompakten Drucksensoren PREMASGARD ® 1110 (Serie) sind mit acht umschalt baren Messbereichen und optional mit Display ausgestattet (acht Geräte in einem) und dienen zur Messung von Über-, Unter- oder Differenz drücken der Luft. Das piezoresistive Mess element ist temperaturkompensiert und garantiert eine hohe Zuverlässigkeit und Genauigkeit.
  • Page 4 ® PREMASGARD 1110 Rev. 2016 - V11 DE Umrechnungstabelle für Druckwerte: Einheit = mbar 1 Pa 0,00001 bar 0,01 mbar 1 Pa 0,001 kPa 0,000101971 mWs 1 kPa 0,01 bar 10 mbar 1000 Pa 1 kPa 0,101971 mWs 1 bar 1 bar 1000 mbar 100000 Pa...
  • Page 5 Bürdendiagramm ® PREMASGARD 1110 Rev. 2016 - V11 DE Bürdendiagramm PREMASGARD ® 1110 PREMASGARD ® 1110 Montageschema 4...20 mA Kompaktform Bürdendiagramm zulässiger Bereich zulässiger 110 Pa 110 Pa 110 Pa Bereich 35 36 40 Betriebsspannung (DC/V) 35 36 40 ÜBERWACHUNGSARTEN: Betriebsspannung (DC/V) Unterdruck: P1 (+) wird nicht angeschlossen, ist luftseitig offen gegen Atmosphäre...
  • Page 6 ® PREMASGARD 1110 Rev. 2016 - V11 DE ® ® Schaltbild* PREMASGARD 1110 Schaltbild PREMASGARD 1110 Ausgang: mit ⁄ ohne Display Ausgang: mit ⁄ ohne Display 4...20 mA 0 -10 V Taster Taster Nullpunkt setzen Nullpunkt setzen min. max. min. max. (auto zero) (auto zero) 1 2 3 4 5 6...
  • Page 7 Konfiguration und Nullpunktkalibrierung Folgende Konfigurationen können über den Die Offseteinstellung erfolgt über ein Potentiometer auf der Leiterplatte DIP-Schalter vorgenommen werden. mit einem Einstellbereich von ± 10 % des Gerätedruckbereiches. Bei Auslieferung steht das Potentiometer in Mittelstellung und ist verlackt. Die Betätigungsschieber sind ohne Werkzeug Alle Einstellungen und Justagevorgänge, die zuvor durchgeführt werden, bedienbar.
  • Page 8 PREMASGARD ® 1111 PREMASGARD ® 1112 (MB: 0...+ xx Pa) (MB: 0...+ xx Pa) Nach erfolgreicher Nullpunktkalibrierung beträgt die Nach erfolgreicher Nullpunktkalibrierung beträgt die Ausgangsspannung (bei Mittelstellung des Offsetreglers) Ausgangsspannung (bei Mittelstellung des Offsetreglers) 0 V bei O Pa Druckdifferenz! 4 mA bei O Pa Druckdifferenz! Ausgangsspannung 0 ...10 V Ausgangsstrom 4...20 mA...
  • Page 9: Wichtige Hinweise

    Wichtige Hinweise Die Einbaulage ist beliebig. Der Spannungsausgang ist kurzschlussfest, ein Anlegen einer Überspannung am Spannungsausgang zerstört das Gerät. Die Druckbereiche (Messbereiche) sind auf dem Geräteetikett angeben. Bei Messdrücken außerhalb dieses Bereiches kommt es zu Fehlmessungen, zu erhöhten Abweichungen oder es kann zur Zerstörung des Druckmessumformer führen. –...
  • Page 10: Technical Data

    ® PREMASGARD 1110 Rev. 2016 - V11 GB Quality product for HVAC sector, accuracy ± 3 % PREMASGARD ® 1110 series calibratable compact pressure sensors are equipped with eight switchable measuring ranges and with or without optional display (eight devices in one) and are used for measuring above-atmospheric, below-atmospheric, or differential pressures in air. The piezo- resistive measuring element is temperature-compensated and guarantees a high degree of reliability and accuracy.
  • Page 11 ® PREMASGARD 1110 Rev. 2016 - V11 GB Conversion table for pressure values: Unit = mbar 1 Pa 0.00001 bar 0.01 mbar 1 Pa 0.001 kPa 0.000101971 mH 1 kPa 0.01 bar 10 mbar 1000 Pa 1 kPa 0.101971 mH 1 bar 1 bar 1000 mbar...
  • Page 12 ® PREMASGARD 1110 Rev. 2016 - V11 GB Load resistance diagram Load resistance diagram PREMASGARD ® 1110 PREMASGARD ® 1110 Mounting diagram 4...20 mA Compact form Load resistance diagram Admissible range Admissible 110 Pa 110 Pa 110 Pa range 35 36 40 Operating voltage (V DC) 35 36 40 TYPES OF MONITORING:...
  • Page 13 ® PREMASGARD 1110 Rev. 2016 - V11 GB ® ® Schematic diagram* PREMASGARD 1110 Schematic diagram PREMASGARD 1110 Output: with ⁄ without display Output: with ⁄ without display 4...20 mA 0 -10 V Pushbutton Pushbutton Zero point setting Zero point setting min.
  • Page 14 Configuration and zero point calibration The following configurations can be preset via Offset adjustment is done at a potentiometer on the circuit board within an adjustment range of ± 10 % of the device’s pressure range. DIP switches. Ex-factory this potentiometer is in center position and is lacquer sealed. The DIP switch sliding blocks can be moved without using tools.
  • Page 15 PREMASGARD ® 1111 PREMASGARD ® 1112 (Range: 0...+ xx Pa) (Range: 0...+ xx Pa) After successful zero point calibration, the output voltage After successful zero point calibration, the output current is 0 V at 0 Pa pressure difference (with the offset knob is 4 mA at 0 Pa pressure difference (with the offset knob in central position)! in central position)!
  • Page 16 General notes This device can be mounted in any position. The voltage output is short-circuit proof. Applying overvoltage at the voltage output will destroy this device. Pressure ranges are indicated on the device label. Applying measuring pressures beyond that range will cause mismeasurements and increased deviations or may destroy the device.
  • Page 17: Caractéristiques Techniques

    ® PREMASGARD 1110 Rev. 2016 - V11 FR Produit de qualité destiné au domaine CVC, précision ± 3 % Les capteurs PREMASGARD ® 1110 (série), capteurs de pression compacts étalonnables, disposent de huit plages de mesure commutables (8 appareils en un) et d’un afficheur en option. Ils servent à la mesure de surpressions, dépressions ou pressions différentielles dans l’air. Grâce à la compensation en température, l’élément de mesure piézorésistif garantit une grande fiabilité...
  • Page 18 ® PREMASGARD 1110 Rev. 2016 - V11 FR Tableau de conversion pour valeurs de pression : Unité = mbar 1 Pa 0,00001 bar 0,01 mbar 1 Pa 0,001 kPa 0,000101971 mWs 1 kPa 0,01 bar 10 mbar 1000 Pa 1 kPa 0,101971 mWs 1 bar 1 bar...
  • Page 19 ® PREMASGARD 1110 Rev. 2016 - V11 FR Load resistance diagram Diagramme de charge PREMASGARD ® 1110 PREMASGARD ® 1110 Schéma de montage 4...20 mA forme compacte Load resistance diagram Admissible range Admissible 110 Pa 110 Pa 110 Pa range 35 36 40 Operating voltage (V DC) 35 36 40...
  • Page 20 ® PREMASGARD 1110 Rev. 2016 - V11 FR ® ® Schéma de raccordement* PREMASGARD 1110 Schéma de raccordement PREMASGARD 1110 Sortie: avec ⁄ sans écran Sortie: avec ⁄ sans écran 4...20 mA 0 -10 V Pushbutton Pushbutton Zero point setting Zero point setting min.
  • Page 21 Configuration et étalonnage du point zéro Les configurations suivantes peuvent être Le point zéro se règle par l’intermédiaire d’un potentiomètre sur la plaquette, effectuées au moyen de l’interrupteur DIP. selon une plage de réglage de ± 10 % de la plage de pression de l’appareil. Les curseurs ne nécessitent aucun outillage.
  • Page 22 PREMASGARD ® 1111 PREMASGARD ® 1112 (plage de mesure: 0...+ xx Pa) (plage de mesure: 0...+ xx Pa) Une fois le calibrage du point zéro effectué, la tension Une fois le calibrage du point zéro effectué, le courant de sortie (si le bouton offset est en position médiane) de sortie (si le bouton offset est en position médiane) est de 0 V pour une différence de pression de 0 Pa ! est de 4 mA pour une différence de pression de 0 Pa !
  • Page 23: Généralités

    Généralités La position de montage peut être choisie librement. La sortie en tension est isolée de la masse, l’application d’une surtension à la sortie en tension causerait la destruction de l’appareil. Les plages de pression (plages de mesure) sont indiquées sur l’étiquette signalétique de l’appareil. Le non-respect des dites plages de mesure entraînera des mesures erronées, des incertitudes de mesure plus élevées ou peut causer la destruction du convertisseur de pression.
  • Page 24: Технические Данные

    ® PREMASGARD 1110 Rev. 2016 - V11 RU Высококачественный датчик для систем отопления, вентиляции и кондиционирования воздуха, точность ±3 % ® Калибруемые компактные датчики PREMASGARD 1110 (серия) с восьмью переключаемыми диапазонами измерения и опционально с дисплеем (восемь приборов в одном) служат для измерения избыточного давления, разрежения и разности давлений в воздухе. Пьезорезистивный измери- тельный...
  • Page 25 ® PREMASGARD 1110 Rev. 2016 - V11 RU Таблица пересчета значений давления: Единицы = бар мбар Па кПа м вод. ст. 1 Па 0,00001 бар 0,01 мбар 1 Па 0,001 кПа 0,000101971 м вод. ст. 1 кПа 0,01 бар 10 мбар 1000 Па...
  • Page 26 ® PREMASGARD 1110 Rev. 2016 - V11 RU Load resistance diagram Нагрузочная диаграмма PREMASGARD ® 1110 PREMASGARD ® 1110 Схема монтажа 4...20 мA Компактное исполнени Load resistance diagram Admissible range Admissible 110 Pa 110 Pa 110 Pa range 35 36 40 Operating voltage (V DC) 35 36 40 ВИДЫ...
  • Page 27 ® PREMASGARD 1110 Rev. 2016 - V11 RU ® ® Схема подключения* PREMASGARD 1110 Схема подключения PREMASGARD 1110 Выход: с дисплеем ⁄ без дисплея Выход: с дисплеем ⁄ без дисплея 4...20 мA 0 -10 В Pushbutton Pushbutton Zero point setting Zero point setting min.
  • Page 28 Конфигурация и калибровка нулевой точки При помощи DIP-переключателей можно Смещение настраивается при помощи потенциометра на печатной плате, выполнить следующие конфигурации. диапазон настройки составляет ± 10 % измеряемого прибором диапазона давлений. Управляющие механизмы приводятся в действие без специального инструмента. При поставке потенциометр находится в среднем положении и залит лаком. Переключатель...
  • Page 29 PREMASGARD ® 1111 PREMASGARD ® 1112 (диапазон измерения: 0...+ xx Па) (диапазон измерения: 0...+ xx Па) После успешной калибровки положения нуля выходное После успешной калибровки положения нуля выходной ток напряжение (в среднем положении регулятора смещения) (в среднем положении регулятора смещения) составляет составляет...
  • Page 30: Напряжение Питания

    Указания к продуктам Установочная длина может быть любой. Выход напряжения защищен от короткого замыкания, приложение завышенного напряжения к выходу напряжения выводит прибор из строя. Диапазоны давлений (диапазоны измерения) указаны на этикетке/табличке прибора. Попытка измерения давления вне этих пределов ведет к погрешностям, повышенным отклонениям или может стать причиной выхода из строя измерительного преобразователя давления. –...
  • Page 31 Reprints, in part or in total, are only permitted with the approval of S+S Regeltechnik GmbH. La reproduction des textes même partielle est uniquement autorisée après accord de la société S+S Regeltechnik GmbH. Перепечатка, в том числе в сокращенном виде, разрешается лишь с согласия S+S Regeltechnik GmbH.