Conçu pour améliorer de manière générale la circulation sanguine, la fonction cardiaque, l'apport en nutriments et en oxygène et l'élimination des produits métaboliques, améliorant par le fait même la qualité de vie de l'utilisateur
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Electromagnetic immunity
Portable and mobile radio equipment should be used no closer to the BEMER SET CLASSIC/PRO and its power cables
than the recommended safety distance calculated from the equation applicable to the frequency transmitter.
EMISSION TESTS
IEC 60601 TEST LEVEL
Conducted RF interference
3 V
eff
according to IEC 61000-4-6
150 kHz to 80 MHz outside of the
ISM bands
Radiated RF interference according
10 V/m
to IEC 61000-4-3
80 MHz to 2.7 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations.
The propagation of electromagnetic waves is affected by the absorption and reflection from structures, objects and people.
Use of BEMER Professional or BEMER Classic products near HF surgical equipment, trains, boats, planes, short wave therapy equip-
ment or inside the RF shielded room of MRI systems may cause degradation or loss of device functionality due to EM disturbances.
72
COMPLIANCE LEVEL
3 V
eff
10 V/m
ELECTROMAGNETIC ENVIRONMENT – GUIDELINES
Recommended safety distance:
d=1.2 P
d=1.2 P
for 80 MHz to 800 MHz
d=2.3 P
for 800 MHz to 2.7 GHz
Where P is the output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d
is the recommended safety distance in meters (m). The field strength from a stationary radio transmitter, according
to an onsite survey
a
is less than the compliance level
b
for all frequency ranges.
Interference may occur in the vicinity of equipment marked with the following symbol.
a) Field strength from fixed transmitters such as base stations for radio (cellular/wireless) telephones and land mobile radios, amateur radio,
AM/FM radio stations and TV stations cannot be predicted theoretically with accuracy. To determine the electromagnetic environment
of a stationary transmitter, a site survey should be considered. If the measured field strength in the location in which the BEMER SET
CLASSIC/PRO is used exceeds the conformity level listed above, the BEMER SET CLASSIC/PRO should be monitored to verify normal
operation. If abnormal performance is observed, additional measures might be necessary, such as readjusting the BEMER SET CLASSIC/PRO
or changing to another location.
b) The field strength is less than 3 V/m over the frequency range of 150 kHz to 80 MHz.