Bemer Classic-Set Mode D'emploi page 35

Conçu pour améliorer de manière générale la circulation sanguine, la fonction cardiaque, l'apport en nutriments et en oxygène et l'élimination des produits métaboliques, améliorant par le fait même la qualité de vie de l'utilisateur
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Guideline and manufacturer's declaration of conformity (Tables according to IEC 60601-1-2)
Electromagnetic emission
The BEMER SET CLASSIC/PRO is designed for operation in the electromagnetic environment listed below.
The customer or user of the BEMER SET CLASSIC/PRO should ensure that the product be used in such an environment.
EMISSION TESTS
COMPLIANCE
RF emisions according
Group 1
to CISPR 11
RF emisions according
Class B
to CISPR 11
Harmonics according
Class A
to IEC 61000-3-2
Voltage fluctuations/flicker
Compliant
according to IEC 61000-3-3
Table according to IEC 60601-1-2
68
ELECTROMAGNETIC ENVIRONMENT – GUIDELINES
The BEMER SET CLASSIC/PRO uses RF energy exclusively for its internal
operation. Therefore, its RF emission is very low. It is unlikely that nearby
electronic devices will be disturbed.
The BEMER SET CLASSIC/PRO is designed for use in all premises, includ-
ing residential areas that are directly connected to a public power supply
that also supplies residential buildings.
Recommended safety distance between portable and mobile RF telecommunication devices and the BEMER CLASSIC/PRO SETS
The BEMER SET CLASSIC/PRO is designed for operation in an electromagnetic environment in which RF disturbances are controlled.
The customer or user of the BEMER CLASSIC/PRO SETS can help to prevent electromagnetic disturbances by keeping a minimum dis-
tance between portable and mobile RF telecommunication devices (transmitters) and the BEMER CLASSIC/PRO SETS – depending on
the power output of the communication device, as indicated below.
OUTPUT POWER
SAFETY DISTANCE, DEPENDENT ON THE TRANSMITTED FREQUENCY m (IN METERS)
OF THE TRANSMITTER W
150 KHZ TO 80 MHZ
d=1.2 P
0. 1
0.38
1
1.2
10
3.8
100
12
For transmitters whose maximum output power rating is not listed in the table above, the distance can be determined by calcu-
lating the equation in each column, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the BEMER Professional and/or BEMER Classic product, including cables speci-
fied by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
fr
es
en
80 MHZ TO 800 MHZ
800 MHZ TO 2.7 GHZ
d=1.2 P
d=2.3 P
0.38
0.73
1.2
2.3
3.8
7.3
12
23

Hide quick links:

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Pro-set410100410200