Conductivity Quick Setup; Measurement Units - Mettler Toledo M400 Guide Rapide

Masquer les pouces Voir aussi pour M400:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 51
Cond Ind One point Slope Calibration
Start as in chapter COND IND ZERO POINT CALIBRATION described. Select the kind of calib r ation.
217.4
µS/cm
25.0
A
°C
Conductivity Calibration
Type = 1 Point Slope
217.4
µS/cm
25.0
A
°C
A Point 1 = 215.0 µS/cm
A C = 217.4 µS/cm

11 Conductivity Quick Setup

For Display setup, analog outputs and setpoints see the section "General Setup".
Sensor Type Selection
1.25
µS/cm
25.0
°C
Sensor Type = Cond(2) m
Cell Constant
1.25
µS/cm
25.0
°C
p M=0.1003 A=0.0000
s M=1.0000 A=0.0000

Measurement units

1.25
µS/cm
25.0
°C
a S/cm
Analog Output? Yes
m
20
Select 1 POINT SLOPE and press [ENTER].
u
Enter the value of calibration point 1 and then press the [ENTER] key to start calibra-
tion.
The cell multiplier and the adder are displayed. After a successful calibration, the
calibration values are taken over and can be stored in the cal history (Adjust), only
u
stored in the cal history (Calibrate) or the calibration can be aborted (Abort).
(PATH: Menu / Quick Setup)
Select the type of sensor to be used with the M400 transmitter. Choices are "Cond(2)",
used for all 2-Electrode type sensors and "Cond (4)" for all 4-electrode sensors.
Press the [ENTER] key.
Enter the appropriate cell constant(s):from the sensor label or certificate (M) for
2-electrode sensors, leaving (A) at 0.000; or (M) and (A) values for 4-electrode
sensors. Press the [ENTER] key.
Select the measurement (conductivity or temperature) and units for measurement.
If using analog output, select YES. Refer back to section "General Setup" to continue
setup.

Publicité

Table des Matières
loading

Table des Matières