Calibration (Inpro 5000) - Mettler Toledo M400 Guide Rapide

Masquer les pouces Voir aussi pour M400:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 51
21.7
H
%sat
25.0
°C
Calibration Successful
8
CO

Calibration (InPro 5000)

2
For Display setup, analog outputs and setpoints see the section "General Setup".
CO
One point Calibration
2
180.4
A
hPa
26.1
A
°C
Calibrate Sensor
Channel A CO2
180.4
H
hPa
26.1
A
°C
CO2 Calibration
Type = 1 Point
137.5
H
hPa
26.1
A
°C
Press ENTER when
Sensor is in Buffer 1
154.5
hPa
26.1
A
°C
A Point1 = 7.00 pH
A CO2 = 7.07 pH
154.5
H
hPa
26.1
A
°C
pH S=100.0% Z=7.048pH
Save Adjust
14
Select ADJUST to save the calibration values and the successful calibration is con-
firmed on the display. If you select CALIBRATE, the values will not be taken. If an ISM
Sensor is used, the Calibration data will be stored in the calibration history.
While in measurement mode press the key [CAL]. Press the m or . key to select the
CO
CALIBRATION and press the [ENTER] key to call the sub function.
2
u
Select 1 POINT and press [ENTER].
u
Place the sensor in the buffer solution and press the [ENTER] key to start the calibra-
tion.
u
The display shows the buffer the transmitter has recognized (Point 1) and the mea-
sured value.
u
As soon as the drift conditions have stabilized the display changes to show the slope
calibration factor S and the offset calibration factor Z.
After a successful calibration, the calibration values are taken over (Adjust) or were
aborted (Calibrate or Abort).
u
(PATH: Cal)