Co Hi (High) Calibration (Inpro 5500 I) - Mettler Toledo M400 Guide Rapide

Masquer les pouces Voir aussi pour M400:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 51
9
CO
Hi (High) Calibration (InPro 5500 i)
2
For Display setup, analog outputs and setpoints see the section "General Setup".
One point calibration
189.0
B
hPa
25.0
B
°C
Calibrate Sensor
Channel B CO2 Hi
189.0
B
hPa
25.0
H
°C
CO2 Hi Calibration
Type = 1 Point
189.0
B
hPa
25.0
H
°C
Press ENTER when
Sensor is in Gas (CO2)
189.0
B
hPa
25.0
H
°C
B Point1=1013. hPa .
B CO2=189.0 hPa
189.0
B
hPa
25.0
H
°C
CO2 S=
mV BL=
mV
9.28
253
Save Adjust
While in measurement mode press the key [CAL]. Press the . or m key to select the
CO
Hi and press the [ENTER] key to call the sub function.
2
u
Select "1 point" and press [ENTER].
u
Place the sensor in the calibration gas (e.g. 100% CO
Press [ENTER].
u
Enter the value for Point 1 including a decimal point and units. The value in the
second text line is the value being measured by the transmitter and sensor in the
units selected by the user.
Depending on the Drift control (see User Manual chapter 8.2.3.10 "CO
u
meters"), one of the two following modes is active.
If Drift is auto, as soon as the signal of stabilization criteria have been fulfilled,
transmitter will display the calibration result automatic.
If Drift is Manual, press [ENTER] to proceed as soon as the value is stable, trans m itter
u
will display the calibration result after manual press [ENTER]
If "Adjust" or "Calibrate" are chosen, the message "Calibration successful" is
d isplayed. In any case you will get the message "Re-install sensor" and "Press
ENTER" on the display. After pressing "ENTER" the M400 returns to the measuring
mode.
(PATH: Cal)
) resp. solution.
2
Hi Para-
2
17