Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
• This device is a Class A device. In a residential
environment, it may cause radio frequency interference (RFI). In this case, the
user may be required to take appropriate measures.
3- Precautions
• Using accessories and cables other than those specified and sold by Integra
MicroFrance can cause an increase in emissions and a decrease in the
device's immunity.
• Check that all the components work before using them to ensure that there is
no damage to the device.
• In the event of any visible damage, do not use this device.
• Ensure that the cables are not tangled. Check the cables and pins to ensure
that they are not split, torn or corroded.
• Use only the recommended cleaning agents.
• Do not sterilize the pneumatic pedal.
• Do not allow the equipment to become humid. If liquid reaches the inside of
the device, damage may occur.
• Store the system in a clean, dry area.
• Keep fingers away from the interruption system. There is a risk of crushing.
• Do not block the vents, as this may prevent the device from cooling properly.
• Connect the equipotential terminal to the surgical terminal.
• This medical device is designed to operate at a maximum temperature of
26 °C and the maximum amount of humidity tolerated is 60%.
• This device is not explosion-proof. Do not use in an explosive atmosphere.
• Position the device in such a way that the mains connector is easily
accessible.
• The device is separated from the power grid by disconnecting the power cord
from the socket behind the device.
4- Adverse events
None known.

II- Indications

This device is designed for irrigation purposes during ENT surgery.
NT010
K – 2015/11
Page 33 sur 91

Publicité

Table des Matières
loading

Table des Matières