Baumer CleverLevel LBFS Manuel De Mise En Service page 7

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5.2 Installation for hygiene applications
WARNING
Danger to health from contaminated medium
Only use weld-in sleeves or adapters from Baumer.
f
Do not seal process connections with Teflon tape (PTFE) or elastomer.
f
Welding work must only be carried out by welders trained in the area of hygiene.
f
1
2
www.baumer.com
Level measurement
CleverLevel
Point level detection
LBFS with the following process connections:
M18x1 ISO 965 (BCID M11)
Tighten the nuts on both sides.
f
Tightening torque: 15...20 Nm
LBFS with the following process connection:
G 1/2 A hygienic (BCID A03)
9 The weld-in sleeve or adapter must be
hygienically mounted and internally flush.
9 Weld seams are smoothed out to Ra < 0.8 µm.
9 Leakage hole points downwards.
Screw in sensor.
f
Tightening torque: 10...15 Nm
LBFS with the following process connection:
G 1/2 A hygienic with adjustable connection
(BCID A03)
9 The weld-in sleeve or adapter must be
hygienically mounted and internally flush.
9 Weld seams are smoothed out to Ra < 0.8 µm.
9 Leakage hole points downwards.
9 The clamping rings must be in faultless condition
(if they are deformed, replace both clamping
rings).
Push the wide clamping ring on to the guide
f
tube.
Push the narrow clamping ring on to the guide
f
tube.
Position the sensor.
f
Adjust the dip depth.
f
Projecting length: 15...228 mm
Tighten the screw-in pin.
f
Tightening torque: 10...15 Nm
Operating Instructions
®
LBFS
7 / 48

Publicité

Table des Matières
loading

Table des Matières