Varian Turbo-V 150 Notice De Mode D'emploi page 62

Table des Matières

Publicité

Character
ASCII
"A"
41h
CRC
-
"B"
42h
CRC
-
"C"
43h
CRC
-
"D"
44h
CRC
-
"E"
45h
CRC
-
"F"
46h
CRC
-
"G"
47h
CRC
-
"H"
48h
"I "
49h
CRC
-
"J"
4Ah
CRC
-
"K"
4Bh
CRC
-
"ACK"
06h
CRC
-
"NACK"
15h
CRC
-
Answer = after a request from the host, the
Turbo-V 150 controller will answer in one of the following
ways:
ACK
NACK
Message
When the Request is "A", "B", "C", "D", "F" the Turbo-V
150 controller will Answer the ACK or NACK.
Converted ASCII
for CRC calculation
-
BFh
-
BEh
-
BDh
-
BCh
-
BBh
-
BAh
-
89h
-
-
B7h
-
B6h
-
B5h
-
FAh
-
EBh
When the Request is "E, the Answer will contain the
complete set of the following parameters:
Bytes
0000=STOP
0001=WAITING INTERLOCK
LSB X
0010= STARTING
X
0011 = NORMAL OPERATION
X
0100= HIGH LOAD
X
0101= ""
1
0110 =FAILURE
0111=APPROACHING LOW SPEED
X
...
X
...
X
...
MSB X
...
Cycle Time
2÷5
6÷9
Pump Life
Pump Temperature
10÷11
12
Current*
13
Voltage*
Frequency
14÷17
Cycle #
18÷19
20
R1 Status
21
R2 Status
22
CRC
*The values for current and voltage are given as numbers,
scaled from 0 to 255, where 0 corresponds to 0 V and 255
to the full scale voltage (130V) or current (5A).
When the Request is "G", the Answer will contain a string
of 11 characters with the following parameters:
Bytes
1-2
Pump cycles number (integer coded
in 2 bytes)
3
Speed threshold
4-7
Run up time in seconds
(long coded in 4 bytes)
8
Deat time
(0 = NO 1 = YES)
9
Soft Start mode
(0 = NO 1=YES)
10
Water cooling
(0 = NO 1=YES)
11
CRC
When the configuration parameters have to be changed,
send a Request string with the letter "H" and a series of
characters as shown below:
"H" + .... data .... + CRC
57
TECHNICAL INFORMATION
Meaning
""
Meaning
87-900-907-01 (B)

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-9436969-9536

Table des Matières