Connection P6 - Low Power Fan; Controller-To-Pump Connection - Varian Turbo-V 150 Notice De Mode D'emploi

Table des Matières

Publicité

Connection P6 - Low Power Fan

P6 connector
The 24 Vdc, 100 mA maximum output voltage is
present on connector P6 after START pushbutton is
pressed and will remain present until a fault condition is
displayed on front panel display or the turbopump is
stopped.
This
is
a
dedicated
Turbo-V 150 pump cooling fan.
The mating connector and cable are provided
assembled with the air cooling kit.
output
for
the
optional

Controller-to-Pump Connection

A five-meter long cable is provided to connect the
controller to the pump.
The following figures show the controller output
connector configuration where pins:
A-F = pump temperature sensor
B-C-D = 54 Vac 3-phase output to pump motor
stator
E = ground
Controller-to-pump connector
54
TECHNICAL INFORMATION
87-900-907-01 (B)

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-9436969-9536

Table des Matières