Stylus Level Setting - Renishaw TS27R Manuel D'installation Et D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

1-12
Square stylus
H, L1 and L2
4 mm AF
5 Nm
(3.69 lbf.ft)
STYLUS TYPES
Disc stylus
Ø12.7 mm (Ø0.5 in)
Square stylus 19.05 mm × 19.05 mm
(0.75 in × 0.7 5 in).

STYLUS LEVEL SETTING

The top surface of the stylus must be set level, front
to back and side to side.
The front to back stylus level is obtained by
alternately adjusting screws L1 and L2, which
STYLUS LEVEL SETTING
TAKE CARE not to stress the break stem
Disc stylus
L1
Screw with
spring washer
L2
L3 and L4
2.5 mm AF
0.8 Nm (0.6 lbf.ft)
L3
causes the probe cable end to rise or lower,
changing the stylus level setting. When a level
stylus surface is obtained, tighten screws L1 and
L2.
Side to side level is obtained by alternately
adjusting grubscrews L3 and L4, which causes the
probe module to rotate and change the stylus level
setting. When a level stylus surface is obtained,
tighten screws L3 and L4.
L4
Side
Back
Front
Side
H
H

Publicité