Support bar
2 mm AF
1.1 Nm
(0.81 lbf.ft)
STYLUS AND HOLDER
Place the stylus and holder onto the break stem and
loosely fit grubscrew B.
Fit screw C through the captive link and into the
stylus holder, then tighten all the screws.
REPLACING THE BREAK STEM
Remove the broken parts and rebuild in the
sequence shown above.
Always hold the support bar in position to counteract twisting
forces and avoid over-stressing the stylus break stem.
B

BREAK STEM

C
3 mm AF
2.6 Nm
(1.92 lbf.ft)
NOTE
Earlier versions of the TS27R probe were supplied
with cup-pointed grubscrews and a different break
stem. Only fit current components supplied with this
probe or with the break stem retrofit kit.