Sanitas SFT 77 Mode D'emploi page 68

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Table 3
Guidance and manufacturer's declaration – electromagnetic immunity
The SFT77 Multi-function Thermometer is intended for use in the electromagnetic environment specified below. The customer or the
user of the SFT77 Multi-function Thermometer should assure that it is used in such an environment.
IMMUnITy
IEC 60601
test
test level
Conducted RF
3Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications equipment should be used no closer to any
part of the SFT77 Multi-function Thermometer, including cables, than the recommen-
ded separation distance calculated from the equation applicable to the frequency of the
transmitter.
recommended separation distance:
d = 1.2
3Vrms
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2.5 GHz
3 V/m
Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site sur-
vey,
a
should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
68
b

Publicité

Table des Matières
loading

Table des Matières