GUIDANCE AND MANUFACTURER'S DECLARATION

Electromagnetic emissions
The « KINETEC Spectra or Spectra DC » is intended for use in the electromagnetic environment specified below. The
customer or the user of the « KINETEC Spectra or Spectra DC » should ensure that it is used in such an environment.
Emissions test
Radio electric-Frequency emissions
CISPR 11
Radio electric-Frequency emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / Flicker emissions
IEC 61000-3-3
Electromagnetic immunity
The « KINETEC Spectra or Spectra DC » is intended for use in the electromagnetic environment specified below. The
customer or the user of the « KINETEC Spectra or Spectra DC » should ensure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient
/ burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage interruptions
CEI 61000-4-11
Voltage dips and
voltage variations on
power supply input
lines
CEI 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE : U
is the a.c. mains voltage prior to application of the test level.
T
Spectra & Spectra DC
Compliance
Group 1
Class B
Class A
Complies
IEC 60601
Test level
±2 kV, ±4 kV, ±6 kV
contact
±2 kV, ±4 kV, ±8 kV
air
±2 kV for power supply lines
±1 kV between lines
±2 kV between line and
earth
< 5% U
T
(>95% dip in U
)
T
for 5 seconds
< 5% U
T
(>95% dip in U
)
T
for 0,5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
3A/m
Electromagnetic environment – guidance
The « KINETEC Spectra or Spectra DC » uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment
The « KINETEC Spectra or Spectra DC » is suitable for use in all
establishments including domestic establishments and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Compliance level
±2 kV, ±4 kV, ±6 kV
Floors should be wood, concrete or ceramic tile. If
contact
floors are covered with synthetic material, the
relative humidity should be at least 30%.
±2 kV, ±4 kV, ±8 kV
air
Mains power quality should be that of a typical
±2 kV for power supply
lines
commercial or hospital environment.
±1 kV between lines
Mains power quality should be that of a typical
±2 kV between line
commercial or hospital environment.
and earth
At the interruption, there is a reset of the « KINETEC
< 5% U
T
Spectra or Spectra DC ». After turning on, push
(>95% dip in U
)
T
for 5 seconds
START to begin the session.
< 5% U
T
(>95% dip in U
)
T
Mains power quality should be that of a typical
for 0,5 cycle
commercial or hospital environment. If the user of
the « KINETEC Spectra or Spectra DC » requires
40% U
T
continued
(60% dip in U
)
T
for 5 cycles
interruptions,
« KINETEC Spectra or Spectra DC » be powered
70% U
T
from an uninterruptible power supply or a battery.
(30% dip in U
)
T
for 25 cycles
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
3A/m
typical commercial or hospital environment.
EN
Electromagnetic environment – guidance
operation
during
it
is
recommended
power
supply
that
the
15/16

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Spectra dc

Table des Matières