Technical specifications

Product :
Weight: 12Kg (24 pounds)
Splint dimensions:
95cm (37 inches) x 33cm (13 inches) x 33cm (13 inches)
Angular limits: -10° to 120°
Speed: from 45 to 155° per minute.
Patient height:
Electricity:
Power supply: 100-240 V~
Frequency: 50-60 Hz
Power consumption: 50 VA
Device of type B class I
IP 20.
Fuse T 750mA 250V 6.3x32mm (KINETEC order: 4610007434)
Environment:
- Storage/transport conditions:

- Operating conditions:

Symbols used

TYPE B device
(protection against electric shocks)
Warning or CAUTION
(check accompanying
documentation)
STOP (power off)
ON (power on)
Program access
Warm Up
Session time
Right way up when box is stored
Contains electric and electronic
components; not to throw in the
dustbins of household refuse

Warranty

The KINETEC warranty is strictly limited to the replacement free of charge or repair in the plant of the
component or components found to be defective.
KINETEC guarantees its joint passive mobilization systems for 2 years against all defects of manufacture from
the date of purchase by the consumer.
KINETEC is the only organization able to assess the application of the warranty to its systems.
The warranty will be considered null and void if the device has been used abnormally or under conditions of
use other than those indicated in the user's manual.
The warranty will also be considered null and void in the event of deterioration or an accident due to
negligence, inappropriate surveillance or inappropriate maintenance, or due to transformation of the
equipment or an attempt to repair the equipment.
Spectra & Spectra DC
full leg: 71 to 99 cm (28 to 39 inches)
Tibia: 38 to 53 cm (15 to 23 inches)
Femur: 33 to 46 cm (13 to 18 inches)
Temperature: -40 to 70°C / -40 to 160°F
Relative humidity: up to 90%
Room temperature: 10 to 40°C / 50 to 105°F
Relative humidity: up to 80%.
Speed
Pause
Increase
Decrease
Timer
Modulation
Temperature Limit during storage or
transport
Fragile
Refer to the operating instructions
EN
Flexion limit
Extension limit
Start movement
Stop movement
Force
Bypass
~
Alternative current
Keep dry during storage or transport
14/16

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Spectra dc

Table des Matières