General Advice On Measuring; Problem Report - Kroeplin B Série Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5. General advice on measuring:

- To protect the measuring contacts the gauge should be brought into its measuring position without touching
the test piece. In doing so the caliper control lever must be depressed completely. The same applies to the
end of a measurement.
- Hold the gauge loosely during measurement; the caliper control lever should be released and the fixed
caliper arm must not be weighted.
- Only those values measured within the device's measuring range are correct (see "9. Technical details").
Anything outside this range should not be measured.
- The gauge should be verified before each series of measurements by using a comparison measurement with
a block gauge; if there is a deviation from the specified measurement the device should be reset
(See 0 point adjustment).
- To reduce measuring inaccuracy the gauge should be calibrated in the measuring mode and in the same
position in which measurement of the test object later takes place.

6. Problem report:

ERR 01
ERR 02
ERR 06
ERR 07
ERR 08
BATT
Unknown Input (Input is not supported)
Unknown Input (Input is not supported)
Tolerance limit is out of the acceptable range
OFFSET is out of the acceptable range (see "9. Technical details")
REL-MODE is outside of indicating range
In case display "Batt" is blinking after switching on the device, change of battery is
urgently necessary. (see "7. Power Supply")
23

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

B440B440sB4r40B4r40sB440tB660 ... Afficher tout

Table des Matières