Using humidifiers
With the exception of AirSense 10, AirCurve 10, Lumis, S9 and H5i
integrated with a humidifier is not connected to the mains power supply, the humidifier
will only operate in passive unheated mode despite the warm-up message appearing
on the device. When the device is connected to the mains power supply, it will return
to active heated mode.
AirSense 10, AirCurve 10, Lumis, S9 and H5i devices operate normally in active heated
mode when connected to the battery or to the mains power supply.

Cleaning and maintenance

WARNING
Do not immerse the RPS II in water, and do not use liquids to clean any part of
the product.
1.
Disconnect the battery from the power supply unit and the device. Remove all
cables.
2. Wipe the exterior of the battery with a clean cloth.
Storing
CAUTION
Batteries must be charged to 100% indicated by four green LEDs and switched off
before storing. You must recharge the batteries to 100% again after six months
of storage. When not in use, all Lithium-Ion batteries self-discharge over time. If
not periodically recharged (ie, every six months), the RPS II will eventually self-
discharge to the point that it can no longer be recharged. If this occurs the RPS II
is no longer usable and is unrecoverable.
Store the RPS II in a cool, dry place.
Note: A fully charged battery when left switched on, self-discharges to 0% of full
capacity within four weeks of storage. When switched off, the battery self-discharges
to 0% of full capacity within six months of storage.
Servicing
The RPS II is intended to provide safe and reliable operation when used and
maintained in accordance with the instructions provided by ResMed. There is no
servicing necessary during the lifetime of the battery.
The service life of the RPS II exceeds 500 recharge cycles. After 500 recharge cycles,
the battery will last about 60% of the time from its original condition. A full charge in
an older battery will not last as long as in a new battery. ResMed recommends that the
RPS II be tested for battery duration regularly. As with all electrical equipment, if any
irregularity becomes apparent, you should exercise caution and contact an authorized
ResMed service representative.
Traveling
Consult with your carrier if you intend to take the battery on board with your device.