Télécharger Imprimer la page

Agilent Technologies Turbo-V 1001 Rack Mode D'emploi page 206

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 39
16
Instructions for Use
Error Messages
MESSAGE
FAULT: TOO HIGH
LOAD
FAULT: SHORT
CIRCUIT
OVERVOLTAGE
206/288
DESCRIPTION
In normal operation, the current
drawn by the pump is higher
than programmed.
During normal functioning (after
the start-up phase), a short-
circuit condition has been
detected in the power circuit.
A failure has occurred in the
controller power supply section,
or the controller has received a
disturbance from the mains.
Turbo-V 1001 Rack Controller User Manual / 87-901-033-01
REPAIR ACTION
Check that the pump rotor
is free to rotate.
Press the START push-
button twice to start the
pump.
Check connections and
shortages between pump
and controller. Press the
START push-button twice
to start the pump.
Press the START button
twice to start the pump
again. If the message is
redisplayed, contact
Agilent for maintenance.

Publicité

loading

Ce manuel est également adapté pour:

X3501-64003