13 SpO2 measurements may be adversely affected in the presence of high ambient light. Shield the sensor area
(with a surgical towel, or direct sunlight, for example) if necessary.
14 Excessive patient movement
15 Venous pulsations
16 Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line.
17 The patient has hypotension, severe vasoconstriction, severe anaemia, or hypothermia
18 The patient is in cardiac arrest or in shock.
19 Fingernail polish or false fingernails may cause inaccurate SpO2 readings.
Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device
components, including batteries.
Product Properties
1 Operation of the product is simple and convenient.
2 The product is compact, lightweight and fully portable..
3 Low Power consumption.
4 Low voltage warning will be indicated when battery voltage is too low to allow normal operation of the
Oximeter.
5 The product will automatically be powered off when inactive for more than 8 seconds.
Product Operation Scope
FINGER PULSE OXIMETER is a portable non-invasive, spot-check, oxygen saturation of arterial haemoglobin (SpO2)
and pulse rate of adult and paediatric patients at home, and hospital (including clinical use in internist/surgery,
Anaesthesia, intensive care and etc). It is not for continuously monitoring.
The FINGER PULSE OXIMETER requires no routine calibration or maintenance other than replacement of batteries.