Replacing the crucible holder
For the use of the oneway crucible C14
Take the cover part and until the ignition
Installation of the oneway crucible C14
C5010.4
24
the decomposition vessels C 7010
and C 7012 must be put in the
in the foundation C5010.4.
These foundation C 5010.4
replaces the normal crucible holding.
wire on the crucible holder like
discribed before.
Screw down the cru-
cible holder from the
shaft and replace it
with the foundation
C 5010.4.
The oneway crucible C14 must
be put into the crucible holding
of the foundation C 5010.4 so
that the ignition wire touches
the oneway crucible C14.
It would the best to stick it
under the ignition wire.
C14

Technical data

Operating pressure PS
Test pressure PT
Volume
Nuclear temperatures
Operating pressures
Sample decomposition
Oxygen valve
(Vacuum operation not possible)
Crucible holder
Detection temperature
Material in contact with the product
Lower part
Threadet joint
Cover
Insert
max. 230 bar (30MPa)
330 bar (33 MPa)
210 ml
≥ 1300 °C
max. 50°C
isochor
non-returnable valve
as in DIN 51 900
Pt 100 integratet
3.4365
3.4365
C7010
C7012
1.4439
2.4610
1.4439
2.4610