Error messages

The following error messages may be displayed (Fig.5) and have
the causes listed below:
Er 1 Sensor not connected or defective (open circuit).
Er 5 Sensor not in media (no change in temperature of the
media after heater active for 3 minutes).
Er 6 Sensor short circuit.
Er 7 Maximum temperature for the media exceeded
(HI-TEMP setting).
Er 8 Calibration error (re-calibration by field service required).

Accessories

H 38
Retaining bar
H 44
Retaining clip
H 50
Twin-jack adapter
H 52
Power supply kit
H 62
Standard sensor 270 mm stainless steel
H 66
Glass sensor
H 70
Extension cable, 1 m long
AD-C1
Adapter (for Corning magnetic stirrer)
12
270 mm glass

Technical data

Measuring range:
°C
-50 ... 450
Resolution:
K
0.1)
Measuring accuracy:
K
±0,2 + sensor tolerance PT 1000
DIN IEC 751 Class A
Setting accuracy:
K
0,1
Type of control:
PID
Control deviation:
K
± 0,5
Supply voltage:
V DC 8 ... 16 (Limited Power Supply)
Current consumption:
mA
8 (at 9V)
Ambient temperature:
°C
0 ... 60
Relative humidity:
%
80
Duty cycle:
%
100
Protection type:
IP 44
Cable length:
mm 350
Male connector:
6-pin DIN 45322
Socket:
DIN 12878 Class 2
Dimensions: (W x D x H) mm 80 x 20 x 90 (without sensor)
Weight:
kg
0.2
ETSD5 0306