Kinetec Spectra Essential Mode D'emploi page 19

Masquer les pouces Voir aussi pour Spectra Essential:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
Maintenance
After 2,000 hours of operation, the KINETEC Spectra Essential requires lubrication and maintenance
operations (lubrication of the joints, pointer stops and ball screws).
The need for maintenance is indicated by display of the message SERV. MOTOR when the system is
switched on.
Despite that warning, you can continue to use your KINETEC Spectra Essential by pressing START, but you
should contact your nearest KINETEC technician to have the maintenance operations conducted as
soon as possible.
When the machine is no longer in working condition, please return it to us, together with its accessories,
for destruction.
A motor running time counter is available. Simultaneously press keys
displays shows RESET TIME 215H (this is an example),
this counter can be reset by pressing the key
Troubleshooting
A spare parts list and technical catalog are available to you on
request from your KINETEC distributor.
If, after connecting the power supply cable to the power supply and
switching on the KINETEC Spectra Essential:
• The display does not indicate any information:
- Check that the electrical socket is live using another device.
- Replace the fuse(s) (25) of the connector with fuses of the same type
and calibre: 2 fuses T 750 mA 250V (6.3 x 32)
(KINETEC order: 4610007434).
- If the display still does not indicate any information, contact your
nearest KINETEC technician.
• Your KINETEC Spectra does not work and the display indicates 50 STOP 25 115,
Press START again.
Your KINETEC Spectra Essential still does not function: Contact your nearest KINETEC technician.
• Your KINETEC Spectra Essential does not function and the display indicates:
SERVICE D1: angle measurement function failure,
or SERVICE D2: no movement,
or SERVICE D3: abnormal consumption,
or SERVICE D4: power failure or disconnected motor,
Contact your nearest KINETEC technician if the same message is displayed after having switched the device
off, then on, and started it by pressing START
Cleaning
Before conducting any cleaning operation, SWITCH the unit OFF and disconnect the power supply.
In order to ensure optimal hygiene, you are advised to clean the machine for each new patient.
Cleaning should be carried out in the environmental conditions specified in the "Technical Specifications"
section below.
Use a DISINFECTANT (PROPANOL/ISOPROPANOL or ALDEHYDE-based solution). Spray the disinfectant on the
SURFACES (plastic shells and metal components).
In order to ensure optimal hygiene, you are advised to clean the covers for each new patient. All the
consumables enable hazard-free disposal.
Elimination and recycling
a • Packing: Packing must be separated from the components plastic and paper/cardboard and given to
the specific sites for recycling.
b • KINETEC PATIENT PAD KIT: Clean with disinfectant then to give it to the specific sites for recycling.
c • Unit: It contains electronic components, cables, aluminium, steel and plastic parts. When the splint is not
operational any more, to dismount and separate in groups from materials and to give them to correct
unit of recycling or return the machine to Kinetec for destruction.
Spectra Essential
.
EN
and
, the
25
7/9

Publicité

Table des Matières
loading

Table des Matières