Parameter
Setting values
STAB .RNG
V . ACC .
ACC .*
FAST
V . FAST
ZER ./TAR .
W/O STB .
W/ STAB .*
AUTOZER .
ON*
OFF
WT .UNIT
GRAMS*, KILOGR .,
CARATS, POUNDS,
OUNCES, TROY .
OZ ., HKTAEL,
SNGTAEL,
TWNTAEL,
GRAINS, PENYWT .,
MILLIGR ., PTS ./LB,
CHNTAEL .,
MOMMES, AUSTR .
CT, TOLA, BAHT,
MESGHAL,
NEWTON
DISP .DIG .
ALL*
MINUS 1
CAL ./ADJ .
EXT .CAL .
E .CAL .USR .
REPRO
SELECT
INT .CAL .*
CAL .SEQ .
ADJUST*
CAL .-ADJ .
ON Z/T
ON*
OFF
* Factory setting
Explanation
Sets the stability to "very accurate" .
Sets the stability to "accurate" .
Sets the stability to "fast" .
Sets the stability to "very fast" .
If the button is pressed: The function of the [Zero] or [Tare] button is executed
immediately .
The function of the [Zero] or [Tare] button is only executed after stability is
achieved .
Activates automatic zeroing . The display is automatically set to zero in case of a
deviation of 0 less than (X) .
Deactivates automatic zeroing . Zeroing must be triggered with the [Zero]
button .
− The device displays the weight in the selected unit .
− The availability of units depends on national legislation and is therefore
country-specific .
"Show all decimal places": All decimal places are shown in the display . The
setting change is not available on conformity-assessed devices .
"Last decimal place off": The last decimal place is switched off .
The [Adjust] button starts an external calibration and adjustment process with
the preset calibration weight .
The [Adjust] button starts an external calibration and adjustment process with
the user-defined calibration weight value .
Starts the reproducibility test . Determines the current calibration frequency with
a user weight .
The interface [CAL] is assigned a selection of permitted menu items from "CAL ./
ADJ ."
The [Adjust] button starts an internal calibration and adjustment process .
The adjustment starts automatically after the calibration .
The adjustment must be started or exited manually after calibration with the
[Confirm] button .
Activates the initial taring/zeroing . The device is tared or zeroed after it is
switched on .
Deactivates the initial taring/zeroing: After it is switched on, the device shows
the same value as before it was last switched off .
System Settings