Connection Tubing - Pari LC SPRINT Mode D'emploi

Masquer les pouces Voir aussi pour LC SPRINT:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 109

Connection tubing

1. Preparation
Check the product:
– Expiry date of the cleaning agent/disinfectant
– Processing limits reached?
2. Cleaning and disinfection
Manual
cleaning:
Manual
disinfection:
Mechanical
cleaning with
disinfection:
3. Steam sterilisation
not usable
not applicable
not applicable
Alkaline cleaning agent, e.g., neodisher
MediClean forte (Dr. Weigert) in conjunction with
neutralising agent, e.g., neodisher
(Dr. Weigert)
Equipment:
– Cleaning device and disinfector in conform-
ance with DIN EN ISO 15883, e.g.,
RDG G7836 CD (Miele)
– Special baskets for Miele instrument dish-
washer
– Compressed air source for blowing dry
Vario TD programme or comparable valid
programmes
– 34 –
®
®