Special conditions X
1. The relationship between the maximum allowable ambient or medium
temperature and the temperature class is shown in the following table:
For U i ≤ 16 V, I i ≤ 190 mA, P i ≤ 200 mW.
LF measuring circuit and temperature measuring circuits:
Temperature Class
2. The measuring systems drawings must be followed for installation.
3. The capacitance and inductance of the connecting cable must be taken
into account for the layout.
4. The metal body of the conductivity sensor or the safety weld-in sockets or
the METTLER TOLEDO housing type InFit 76*-*** resp. InTrac 7**-***
must be electrically connected to the main potential equalization system
of the plant (grounded).
5. The metal body of the conductivity sensor or the safety weld-in sockets or
the METTLER TOLEDO housing type InFit 76*-*** resp. InTrac 7**-***
are, if necessary, to be included into the recurring pressure test of the
unit.
InPro 7100-VP Series
52 002 092
Max. ambient resp.
medium temperature
T6
T5
T4
T3
51 °C
63 °C
91 °C
143 °C
© 05 / 10 Mettler-Toledo AG, CH - 8606 Greifensee