Press the PUMP CURRENT push-button and the
display shows
L O W E R
T 2
S E N S O R :
Detection of the pump's lower bearing temperature. OFF
will always be indicated if there is no sensor on this
bearing.
and then:
S O F T
S T A R T
S E L E C T I
where: XXX = YES or NO.
If YES is selected, the Soft Start mode allows the pump to
rump-up the Normal speed within ten steps.
When NO is selected, the Soft Start mode is deselected
and the rump-up of the pump will be done within 180
seconds.
The controller is factory set to YES.
The Soft Start mode may be deselected/selected only
when the pump is stopped.
Press CYCLE NUMBER push-button to select YES or
NO
S O F T
S T A R T
S E L E C T I
87-900-902-01 (G) - TURBO-V 300 ICE CONTROLLER
B E A R I
N G
O F F
M O D E
O N :
X X X
NOTE
M O D E
O N :
Y E S
TECHNICAL INFORMATION
Press PUMP CURRENT push-button and the display
shows:
W A T E R
N O
If necessary press CYCLE NUMBER to select YES or NO.
The controller is factory set to NO.
Enter the selection by pressing the PUMP CURRENT
push-button, and the display shows:
F R O N T
/
R E M O T E
S E L E C T
I O N : X X X X X X
where: XXXXXX = means the word FRONT. REMOTE, or
SER depending on the last selection.
The controller is factory-set for FRONT panel operation.
SER will only be displayed if the optional serial card is
installed.
C O O L I
N G
/
S E R
?