adding a new profile.
Copying an existing profile
Procedure:
1. Enter the main menu, to do it press Setup key.
2. Enter <Profiles> submenu.
3. Press and hold profile that is to be copied.
4. Menu referring to a profile record is displayed. Select <Copy> parameter.
5. A new <Name copy> profile is created, new profile settings and base profile settings are identical.
6. Modify necessary data: (name, etc.).
Adding a new profile
Procedure:
1. Enter main menu, to do it press Setup key.
2. Enter <Profiles> submenu.
3. Press
button, message <Create new record?> is displayed.
4. Press
button to confirm. The software automatically adds a new record, the new record edit
box opens.
Deleting a profile
Procedure:
1. Enter the main menu, to do it press Setup key.
2. Enter <Profiles> submenu.
3. Press and hold profile that is to be deleted.
4. Menu referring to a profile record is displayed. Select <Delete> parameter.
5. Message box: <Delete?> is displayed.
6. Press
button to confirm, the profile gets deleted.
Caution! In order to operate profiles it is necessary to log in as an Administrator.

13.2. Profile Structure

Profile parameters list:
Settings
Working modes
Readout
Units
Enter this menu to set an individual profile name (alphanumeric characters
string), and to declare a default working mode (the selected mode is run as
default upon profile activation).
Mode submenus:
Additional settings of a working mode
Buttons
Information
Printouts
Readout submenus:
Filter
Value release
Autozero
Autozero: Dosing
Last digit
Enter this menu to declare start unit, supplementary unit, 2 custom units and to
enter the value of gravitational acceleration force of the place of use.
- 51 -