Name: press "Name" field, a window for entering series name opens.
Code: enter series code.
Customer: select a customer for whom the series measurements are to be carried out.
Samples: press to view list of samples, for a new series the list is empty, add new samples.
to do it press <
and changed.
Samples quantity – parameter disabled for editing, the number is updated at real time by the
software on adding a new sample to the series.
For each sample on the list, its current differential weighing status is displayed (weighing process
stage).
Newly added samples do not have any status assigned – the field is empty. The status is updated at
real time after each completed weighing process stage.

25.4. Differential Weighing Process Example

Activate Differential Weighing mode, next:
1. Select information to be displayed in the grey workspace.
2. Select appropriate quick access buttons.
3. Add a series to balance's memory (series is identified by name).
4. Add samples to a series (sample is identified by name).
5. Return to the differential weighing home screen.
Selecting a series:
Press <SERIES> quick access button, a window with
series database opens.
Enter the database, press field with series that is to be
carried out.
Selected series name is displayed in the grey workspace
(provided that such information is enabled to be
displayed).
Upon selecting a series, press one of the following buttons determining how the differential weighing
is to be carried out:
Weighing A
Press to start weighing initial mass.
Tare (T)
Press to start sample container taring.
Add> button, the sample is added automatically, its name can be edited
- 110 -