Clamping System - HAVER & BOECKER EML 200 Pure Notice D'utilisation

Tamiseuse de laboratoire
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3.2. Clamping system

Classic
Elevation adjustment:
By turning the nut it moves up and down the guide rod.
Tensioning:
The sieve cover has to be installed and both nuts must be screwed down simultaneously.
4. Technical data for the EML 200 Pure
Test sieve shaker with electromagnetic drive and three-dimensional sieving action for dry and wet sieving.
Operation voltage:
Frequency:
Power consumption:
Control:
Weight:
max. weight of full sieve tower:
Dimensions:
Emitted sound:
© Copyright 2010 by HAVER & BOECKER
1.
Nut with handlebars
2.
Sieve cover
3.
Guide rod
110V - 230V
50 Hertz - 60 Hertz
approx. 490 Volt-ampere with full charge
all-electronic
38 kg, without test sieves
8,7 kg
385 x 295 x 900 mm (Dept x Width x Height)
The emitted sound of the machine in action has been determined with a
comparative sound measurement according to DIN 45635-01 KI1.
The equivalent continuous sound level A is: L
The intensity of sound can be reduced when using HAVER
Sound absorbing Cupboards.
≤ 70 (dB)
pAeq
Page 15/31

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Haver eml 200 pure

Table des Matières