Télécharger Imprimer la page

Oldham C 1100 Manuel Technique page 16

Publicité

Les langues disponibles

Les langues disponibles

Default Push Button Functions
On Mode
Normal operation
Alarm enabled
TWA logger active
ABC algorithm disabled
Power Off
TWA logger data will be
reset.
CALb Mode
TWA logger on hold
ABC algorithm enabled
Back light & Application
Function Execution.
Acknowledge Alarm
Push the button until the display shows "tESt", then release the
The unit will then perform a full self-test sequence
button.
showing the LEDs and buzzer response for the alarm
setpoints
. After the test the unit is in On Mode.
Off Mode
"tESt"
On Mode
From On Mode, push the button until all front panel leds are
turned off. While pushing the button the unit will pass through the
mode "CALb". If you accidentally release the button too early,
just push the button once to reach On Mode and try again.
On Mode
"CALb"
Off Mode
From On Mode, push the button until the display shows CALb,
then release the button. This will activate the Automatic Baseline
Calibration ABC that requires 5 hours to complete one calibration
cycle. The instrument continues to execute calibrations every
four hours as long as it is left in CALb mode. Intended for
overnight charging in fresh air, with minor calibration adjustment
automatically performed.
To get back to On Mode, just push the button once.
On Mode
" CALb"
Can be turned on temporary by pushing the button. The
backlight is turned off after 8-16 seconds. Application function for
"standard" default setting is just backlight.
In alarm and pre alarm states, pushing the button resets the
sounding buzzer for 5 minutes at a time.
Explanation
Push Button
Release Button
Page 16 de 22