Angle adjustment of the head rest
The head rest can be adjusted to
a total of 9 different positions.
Pull the top of the head rest upward. Allow it to lock into place at the
desired angle. To lower, pull the top part of the head rest all the way
up again as far as it can go. Then guide it all the way down until it is
at its lowest position.
Now you can readjust the head rest to another position. Gently guide
the head rest with your hand while adjusting. The padding distance
(in accordance with DIN EN ISO 13857:2008-06) makes it impossible
to pinch your hand or fingers.
Maintenance / Cleaning / Disinfection
The exam tables are functionally maintenance-free. The user must however inspect them every 6 months
for the purpose of ensuring themselves that the tables are in proper working order. Hereby, the screws are
to be all tested to ensure that they are all still tightly fastened! Room temperatures of less than 70 C lead
to rust damage on the chrome-plated parts. Our surfaces can be wiped clean with either a dry or damp
cloth. Afterward, the surface must be wiped dry!
Caution! Do not use a dripping wet cloth as water should not be allowed to enter the structural
parts of the table!
Never use abrasive cleaning agents! These will scratch the surfaces (will also scratch the chrome). The
chrome-plated surfaces may not come into contact with halogens such as fluoride, iodine, bromine, or
chlorine/chloride (hydrochloric acid) or their derivatives, as this will damage the chrome or nickel layer.
Chrome-plated parts cannot replace stainless steel in areas with strict hygiene requirements!
The active ingredients of disinfecting agents known to us can cause surface damage when the wrong product
is selected and when used improperly. For disinfection we recommend the products "Mikrozid sensitive
liquid" and "Mikrozid sensitive wipes" made by Schülke & Mayr*. Special care must be taken to remove
all agents from the surface after their relatively short contact time of one minute, otherwise the resulting
chloride compound can cause surface damage due to corrosion – also on stainless steel! The compound
may in no way be allowed to enter the structural parts of the table as the build-up of the concentration of
the residual active ingredients will lead to damage and failure of the mechanical elements.
For this reason, avoid using spray-on disinfectant at all costs! Please remember that it is necessary to
completely remove all agents after their contact time has expired in order to avoid subsequent damage.
Should these recommendations not be followed, a much shorter lifespan of the exam table can be expected.
Even though we have recommended certain disinfection products, the responsibility of critically examining
the surface disinfection process is in the hands of the user. When in doubt, please contact the manufacturer
of the disinfectant with questions about the suitability of their product as they will be best able to determine
whether or not changes have been made to the active ingredient formula of their product. Please note,
that possible material damage often becomes visible only after a long application period.
Leather imitation padding
The table covers do not require conditioning. Normal impurities can be wiped from the cover with a damp
cloth. The cover is to then be rubbed dry with a towel. Tougher messes can be cleaned with grease-free,
soapy water. Never use solvents or greasy cleaning agents! Protect your covers from skin oils, alcohol-based
solutions and creams – they could make the material brittle!
Be especially careful with disinfecting agents!
*Schülke & Mayr GmbH, Robert-Koch-Str. 2, D-22851 Norderstedt
Fax 49 40 521 00 - 318, www.schuelke.com
Installing the paper roll holder
0-34°
2
Part
On the upper part of the legs at the upper side of the table Part 2 (table - top side), there are two
adjacent holes. Put a roll of paper a onto the paper roll holder b and insert the side without the groove
into the hole (see diagram) on the one leg.
Then insert the side of the paper roll holder b with the groove into the hole on the opposite leg (see
diagram),
whereby care must be taken that the groove, which stops the holder from sliding from side to side, is
properly inserted into the wall of the leg.
Now you may unroll paper as needed for the protection of the patient.
5
a
b
Safety Information
- The table serves as an examination, treatment and/or therapy table.
Any other use than that described here can lead to injury or damage
to this product. The exam table is not to be altered. Prohibited action
voids our warranty and CE certification.
- The head rest may only be subjected to the weight of the head (on
the center of the head rest) and parts of the upper body – the head
piece may absolutely not be sat on or the edges otherwise subjected
to a load greater than 10kg.
- Heed the load bearing capacity of the table!
- Do not use this table (with or without) side rails as a resting or
wake-up table – always use 2 side rails!
- The table may not be equipped with rollers or wheels.
- Repairs may only be made by authorized, qualified personnel.
- In order to protect the patients, please use medical sterile covers on
the upholstery .
- Should you notice abnormal changes/noises/hazards while using
this table, we ask that you inform us immediately. Until the issue
has been resolved, the table is not to be used.
Accessories
- Paper roll holder, chrome-plated, for table width of 65 cm
Replacement parts
Plugs (set of 4), paper roll holder,
headpiece adjustment mechanism, hinges (incl. screws)
Symbol key:
Manufacturer
LOT Date of manufacture
Heed the User's Manual
2
Part
Groove
Caution!
Complies with relevant EU
guidelines