Bure Rise & Go DB
The Bure Rise & Go Double is a conventional, electrically operated Walker that is enhanced with a power rise
function. Not only does it act as a Walker, it also makes getting about in everyday life easier thanks to its smart
power rise function.
The Bure Rise & Go can be used by anyone who regularly uses a Walker. The user can remain active while the
power rise function is in use.
The product consists of a Walker and its associated harness and straps. The harness is available in different sizes
(S, M, L, XL) and is split at the rear to make remaining in the harness possible while using the WC. The harness is
secured to the two attachment straps (one each side), which are in turn attached to the Walker.
When the patient stands up the electric power rise function is employed and the smart attachment reduces the
distance between the patient and the Walker until the patient is standing upright.
The Walker can be used to advantage with a platform to facilitate patient transportation.
1. Intended use
Walkers are indoor, movable and portable assistive aids, which together with approved accessories are intended to
be used for transfer from sitting to standing positions, and walking training. Walkers are to be used by persons with
reduced muscular strength, who shall be able to stand and perform simple leg movements. Walkers are to be used
for leaning on when walking and standing during early mobilization training. Walkers and accessories are intended
for indoor use only in Home Healthcare environment and Professional Healthcare facility environment.
2. Contraindications, Precautions and Warnings
Contraindictions
The device may not be used by patients above the maximum weight indicated on the label of the device.
Precautions
Check that the device is used on flat surface.
Check the device is correctly mounted/assembled before its first use.
Check the device after every folding after any transport.
Check height adjustment and leg spreading motions and inspect the actuators' or gas pistons' full ranges.
Activate the brakes whenever the device is not in use.
Check the function of the brakes before use.
Inspect the device regularly to detect any signs of damage. If there are any visible signs of damage, contact
Direct Healthcare Group Customer Service.
In case of an electrical model of the device, check that the hand control does not show signs of wear before
every use.
Check that the hand control markings are in accordance with the lifting functions.
Check the battery status of an electrically operated model, before using the device.
It is important to never leave the user alone during the transfer.
Warranty applies only if repairs or alterations are made by personnel who are authorized by Direct Healthcare
Group.
Ensure that there are no obstacles or people in the way of the device, when moving.