Télécharger Imprimer la page

Bibby Scientific Merit Water Still W4000 Mode D'emploi page 2

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Merit Water Still
W4000 & W4000/EURO
Thermostat reset
buttons
Heater
Grommet
Figure 1 General assembly
Before use
If the equipment is not used in the manner described in this
manual the protection provided by the equipment may be
impaired.
Note: The Merit Water Still is classified as "Permanently
Connected Equipment" and should be connected to the
electricity supply by a qualified electrician in the manner
described in the electrical installation section of this manual.
The Merit Water Still is designed to operate under the following
conditions:
❖ For indoor use only
❖ Ambient temperature range +5°C to +40°C
❖ Altitude up to 2000m
❖ relative humidity not exceeding 80%
❖ Mains supply voltage fluctuations not
greater than ±10% of nominal
❖ overvoltage category II IEC60364-4-443
❖ Pollution degree 2 IEC664
The Merit Water Still complies with the
following European Directives:
89/336/EEC - E.M.C. DIrECTIVE
Amended by 92/31/EEC and 93/68/EEC
73/23/EEC - L.V.D. Amended by 93/68/EEC
2
Condenser upper outlet
Condenser
Boiler
Gaskets and
Vent
flange
Merit
water still
W4000
Distillate
Straps
outlet
Limitations of use
This product meets the applicable E.C. harmonised standards for
radio frequency interference and may be expected not to interfere
with, or be affected by other equipment with similar qualifications.
We cannot be sure that other equipment used in its vicinity will meet
these standards and we cannot guarantee that interference will not
occur in practice. Where there is a possibility that injury, damage or
loss might occur if equipment malfunctions due to radio frequency
interference, or for general advice before use, please contact the
Service Department of Bibby Scientific Ltd.
English
Condenser lower inlet
Constant level
control
outlet
Stopcock

Publicité

loading

Ce manuel est également adapté pour:

Merit water still w4000/euro