1. inflation pressure is higher than 300 mmHg,
2. blood pressure is unusually high or low,
3. you move or talk during the measurement (in addition to
, the cardiac arrhythmia icon
play),
4. the cuff tube is not inserted properly,
5. the cuff is inflated for longer than 16 seconds.
In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do
not move or talk. Re­insert the batteries if necessary, or else
replace them.
9. Specifications
Model no.
BM 19
Measurement
Oscillometric, non­invasive blood
method
pressure measurement on the upper
arm
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 40 – 280 mmHg,
diastolic 40 – 280 mmHg,
Pulse 40 – 199 beats/minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement inac­
Max. permissible standard deviation
curacy
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
Weight
Cuff size
appears in the dis­
Permissible operating
conditions
Permissible storage
conditions
Power supply
Battery life
Accessories
Classification
Technical information is subject to change without notification
to allow for updates.
• This unit is in line with European Standard EN 60601­1­2
21
L 147 mm x W 104 mm x H 66 mm
Approx. 310 g (without batteries)
23 to 33 cm
+5 °C to +40 °C, 15 % ­ 93% relative
air humidity (non­condensing)
­25 °C to +70 °C, ≤93 % relative air
humidity, 700 –1060 hPa ambient
pressure
4 x 1,5 V
For approx. 250 measurements,
depending on the blood pressure level
and/or pump pressure
Instruction for use, 4 x 1.5 V AA
batteries, storage pouch
Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part
and is subject to particular precautions with regard to elec­
tromagnetic compatibility (EMC). Please note that portable
and mobile HF communication systems may interfere with
this unit. More details can be requested from the stated Cus­
tomer Service address or found at the end of the instructions
for use.